Video Transcription
Video Summary
The issue highlighted is that the pacing stimuli are not effectively capturing all electrograms and QRS complexes, evident in the HIS-1-2 lead. This suggests unreliable capture during VT mapping. The recommended action is to reposition the catheter for reliable capture, allowing for effective site interrogation.
×