(FIT)-Module 5 Workshop 9: Arrhythmia Case Studies ...
Ellenbogen Case 5B Answer
Ellenbogen Case 5B Answer
Back to course
[Please upgrade your browser to play this video content]
Video Transcription
Video Summary
The patient likely experienced bacterial seeding at the time of their ICD generator change, which was exacerbated by local trauma. Due to the risk of pocket infection, which might manifest up to a year after the procedure, the recommended treatment is to extract the entire ICD system and use a wearable external defibrillator until intravenous antibiotics are complete. Pocket infection necessitates complete system removal, as oral antibiotics only delay definitive treatment and could increase mortality. Pocket needle aspiration is contraindicated, as per HRS guidelines. Blood cultures and TEE should also be conducted.
Keywords
ICD generator change
bacterial seeding
pocket infection
wearable external defibrillator
system extraction
×